2-2 Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual Radical-7 Color Display Signal Extraction Pulse Oximeter Operator’s Manual 2-32-3
2
system description
1
HANDHELD
RELEASE
BUTTON
Press down the Handheld Release Button and pull the
Handheld device off the Docking Station.
2
97
SpO
2
MEASUREMENT
DISPLAY
The functional arterial hemoglobin oxygen saturation is
displayed in units of percentage SpO
2
. The upper and lower
SpO
2
alarm limits are also displayed next to the SpO
2
mea-
surement. When a sensor is not connected to a patient and
during pulse search, the display will show dashed lines and
the message Sensor Off will appear at the top of the display
screen. When the measured value is outside of the alarm
limits, the SpO
2
Measurement Display flashes and an alarm
will sound. The oxygen saturation is calculated and the dis-
play is updated at a frequency of once per second.
3
Masimo
rainbow
SET
MASIMO
RAINBOW SET
The Masimo Rainbow SET label is shown on the Radical-7
display when SET processing is active.
NOTE: Patient information, not Masimo SET will be dis-
played when using a RadNet compatible unit connected to a
RadNet central station.
4
99
85
SATURATION
ALARM LIMITS
DISPLAY
The Saturation Alarm Limits Display shows the upper and
lower saturation alarm limits. When an alarm limit is exceed-
ed, the SpO
2
value and the violated limit flashes.
5
ALARM STATUS
INDICATOR
The alarm status indicator (a bell) can be shown with or
without a slash. It flashes when an alarm condition is present.
When the alarm is silenced using the Alarm Silence Button,
an alarm status indicator with a slash and a timer is shown
to indicate that the alarm is temporarily silenced. When the
alarm is silenced through the All Mute menu selection (which is
permanent until power is cycled or deselected using the menu)
an alarm status indicator with a slash is shown to indicate that
alarm has been silenced.
6
1.65
SpMet*
MEASUREMENT
DISPLAY
The measurement of methemoglobin concentration levels
is displayed in units of percentage SpMet. The upper and
lower SpMet alarm limits are also displayed next to the
SpMet measurement. When a sensor is not connected to
a patient and during pulse search, the display will show
dashed lines and the message Sensor Off will appear at
the top of the display screen. The methemoglobin is calcu-
lated and the display is updated at a frequency of once per
second.
7
3.0
0.3
SpMet ALARM
LIMITS DISPLAY
The SpMet Alarm Limits Display shows the upper and lower
alarm limits. When the measured value is outside of the
alarm limits, the SpMet measurement display flashes and
an alarm will sound.
8
Sensor Off
SYSTEM
MESSAGE AREA
The system messages generated by the instrument are dis-
played in the System Message Area. See Section 5, System
messages.
9
20
SpCO*
MEASUREMENT
DISPLAY
The measurement of carbon monoxide concentration levels
is displayed in units of percentage SpCO. When a sensor
is not connected to a patient and during pulse search, the
display will show dashed lines and the message Sensor Off
will appear at the top of the display screen. The carboxy-
hemoglobin is calculated and the display is updated at a
frequency of once per second.
system description
2-2
2
Radical-7 Pulse CO-Oximeter
Handheld
The Handheld Radical-7 Pulse CO-Oximeter provides most of the functionality of the Pulse
CO-Oximeter. All user input and displays are controlled by this part of the Radical-7 Pulse
CO-Oximeter system. The sensor cable connects into the connector on the Handheld unit.
The Handheld is battery powered and can be used either as a transport monitor or as a
Handheld Pulse CO-Oximeter for spot checks.
HANDHELD FRONT PANEL
The following figure and corresponding text outline all the features of the Handheld Radical-
7 Pulse CO-Oximeter:
Pleth + Signal IQ View
Numbers View
Signal Extraction Pulse CO-Oximeter
Masimo
rainbow
SET
Sensor Off
99
85
Sp0
2
140
50
BPM
PI
SpMet
SpCO
3.0
0.3
50
5
FastSat APODAPOD 10/11/06 12:23:11am
APOD
Philips VueLink
1514
13
3
2
23 21
16
22
1
25
4
26
20 18
8
5
11
12
24
17
6
9
19
27
Signal Extraction Pulse CO-Oximeter
Masimo
rainbow
SET
Sensor Off
140
50
BPM
PI
FastSat APOD 10/11/06 12:23:11am
99
85
Sp0
2
SpMet
3.0
0.3
SpCO
50
5
Philips VueLink
2
4
9
27
25
23
21
10
7
6
28