4-2 Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual
4
Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual 4-3
4
operation
An alarm tone sounds.
The violated alarm parameter flashes.
12. Verify the sensor alarms are functional by removing the sensor from the sensor site.
“SEn OFF” message appears on the display.
The alarm tone sounds.
The Visual Alarm Indicator flashes.
Disconnect the sensor from the patient cable or oximeter.
Confirm that “NO SEn” message appears on the display.
Note: “NO SEn” and “SEn OFF” will only generate an alarm if the Rad-57 was
actively monitoring a patient when the sensor was disconnected.
13. Verify parameter violation alarm silence operation.
Create an alarm condition by lowering the SpO
2
or pulse rate high alarm limits
beyond the patient readings.
Press the Alarm Silence button.
The alarm tone ceases for 120 seconds.
14. To begin patient monitoring:
Adjust the alarm limits.
Adjust the alarm volume.
Adjust the pulse beep volume.
15. Verify the sensor is applied correctly and that the measured data is appropriate, see
Section 4, Successful SpO
2
Monitoring.
16. Monitor the patient.
17. After monitoring is complete, remove the sensor from the patient and store or
dispose of the sensor according to governing rules. See the Directions for Use of
the sensor.
18. Press and hold the Power On/Off button for 2 seconds to turn the oximeter off.
Note: turn the oximeter off between patients so that it can re-calibrate in order to interpret
new physiological data and to conserve battery life.
operation
DEFAULT SETTINGS
The Rad-57 stores two types of default values: those that the device automatically reverts
to after a power cycle, and those that can be changed by the user and will be remembered
after a power cycle.
The following table outlines the default values that the Rad-57 reverts to after a power
cycle:
OPTION DEFAULT SETTING
SpO
2
high alarm limit Set to Off
SpO
2
low alarm limit Set to 90%
Pulse rate high alarm limit Set to 140 BPM
Pulse rate low alarm limit Set to 50 BPM
SpCO high alarm limit
Set to pre-power down setting
(Factory Default is 10%)
SpCO low alarm limit
Set to pre-power down setting
(Factory Default is off)
Averaging Time Set to pre-power down setting
FastSat Set to pre-power down setting
Sensitivity Set to pre-power down setting
Display brightness Set to pre-power down setting
Pulse tone volume Set to pre-power down setting
Alarm Volume Set to pre-power down setting
Line Frequency Set to pre-power down setting
Trend Active Set to pre-power down setting
Alarm Silence Set to all alarms active