
1-4 Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual
1
Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual 1-5
1
Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
Pulse Oximetry
SpO
2
GENERAL DESCRIPTION
Pulse oximetry is a continuous and non-invasive method of measuring the level of arterial
oxygen saturation in blood. The measurement is taken by placing a sensor on a patient,
usually on the fingertip for adults, and the hand or foot for neonates. The sensor connects
to the pulse oximetry instrument with a patient cable. The sensor collects signal data from
the patient and sends it to the instrument. The instrument displays the calculated data in
two ways: 1) as a percent value for arterial oxygen saturation (SpO
2
), and 2) as a pulse
rate (PR). The following figure shows the general monitoring setup.
SpCO GENERAL DESCRIPTION
Pulse CO-Oximetry is a continuous and non-invasive method of measuring the levels of
carbon monoxide concentration (SpCO) in arterial blood. It relies on the same principles
of pulse oximetry to make its SpCO measurement. The measurement is taken by placing a
sensor on a patient, usually on the fingertip for adults. The sensor connects directly to the
pulse CO-Oximetry instrument or with a patient cable. The sensor collects signal data from
the patient and sends it to the instrument. The instrument displays the calculated data as
percentage value for the SpCO. The Rad-57 is a combined SpO
2
and SpCO monitor with
the same setup as that of a pulse oximeter, shown above, and can display a percentage
value for SpCO as well as SpO
2
.
PRINCIPLE OF OPERATION
Pulse oximetry is governed by the following principles:
1. Oxyhemoglobin (oxygenated blood), deoxyhemoglobin (non-oxygenated blood) and
carboxyhemoglobin (blood with carbon monoxide content) species differ in their
absorption of visible and infrared light (spectrophotometry, see figure below).
650 700 750 800 850 905 950 1000
Wavelength (nm)
1
2
3
4
0
5
6
7
Absorption (mm-1)
Deoxyhemoglobin
Oxyhemoglobin
Carboxyhemoglobin
overview
1
Rad-57 Signal Extraction Pulse CO-Oximeter Operator’s Manual
1
overview
2. The amount of arterial blood in tissue changes with your pulse (photoplethysography).
Therefore, the amount of light absorbed by the varying quantities of arterial blood
changes as well.
The Rad-57 Pulse CO-Oximeter uses a multi-wavelength sensor to distinguish between
oxygenated blood, deoxygenated blood, and blood with carbon monoxide content. Signal
data is obtained by passing various visible and infrared lights (LED’s, 400 to 1000nm)
through a capillary bed (for example, a fingertip, a hand, a foot) and measuring changes
in light absorption during the blood pulsatile cycle. See figure below. The photodetector
receives the light, converts it into an electronic signal and sends it to the Rad-57 for
calculation.
Once the Rad-57 receives the signal from the sensor, it utilizes Masimo Rainbow SET
signal extraction technology to calculate the patient's functional oxygen saturation,
fractional concentration of carboxyhemoglobin, and pulse rate. The SpCO measurement
relies on a multiwavelength calibration equation to estimate the percentage of carbon-
monoxide in arterial blood.
FUNCTIONAL VS. FRACTIONAL SATURATION
The Rad-57 is calibrated to measure and display functional saturation (SpO
2
): the amount
of oxyhemoglobin expressed as a percentage of the hemoglobin that can transport oxygen.
The Rad-57 does not measure fractional saturation: oxygenated hemoglobin expressed as
a percentage of the four main hemoglobin species, i.e., oxyhemoglobin, deoxyhemoglobin,
carboxyhemoglobin, and methemoglobin. To convert fractional saturation to functional
saturation, the fractional saturation measurements must be converted according to:
Functional saturation =
Fractional saturation
x 100
100 - (% carboxyhemoglobin + % methemoglobin)
MEASURED VS. CALCULATED VALUES
SpO
2
and SpCO measurements obtained from the Rad-57 are commonly compared to
invasive measurements obtained from blood gas samples. When comparing invasive
and noninvasive measurements and interpreting values, caution should be used, as the
calculated values obtained from the blood gas sample may differ from the SpO
2
and
SpCO measurements of the Pulse CO-Oximeter. In the case of SpO
2
, different results
are usually obtained from the arterial blood gas sample if the calculated measurement is
not appropriately corrected for the effects of variables that shift the relationship between
the partial pressure of oxygen (PO
2
) and saturation, such as: pH, temperature, the partial
pressure of carbon dioxide (PCO
2
), 2,3-DPG, and fetal hemoglobin. In the case of SpCO,
in addition to the effects of temperature and pH, different results are also expected if
the oxygen saturation and/or concentration of methemoglobin in the blood gas sample
2
1
1. Light Emitting Diodes (LEDs)
2. Recessed Photo Detector
1
3
2
1. Instrument
2. Sensor
3. Patient Cable
Commenti su questo manuale