8-2 Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual
8
Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual 8-3
8
Masimo Rainbow
TM
sensors
Masimo Rainbow sensors must be used with the Rad-57 to enable measurement of
Carboxyhemoglobin (SpCO). Rainbow sensors will only function with oximeter devices
equipped with Masimo Rainbow SET Technology.
Rainbow sensors connect to the device directly or with a patient cable.
RAINBOW REUSABLE SENSORS
SpO
2
and pulse rate accuracy for the Rainbow sensors is specified in the following table.
Weight Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
SENSOR Range No Motion Motion No Motion Motion Saturation Pulse Rate
Rainbow
DC-I-DC
> 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3%
SpCO accuracy is specifed as ±3% under no motion conditions.
Masimo SpO
2
sensors
The Rad-57 may also use standard Masimo LNOP, LNOPv and LNCS SpO
2
sensors,
when used with Red PC and LNCS Cabels respectively. This will allow the Rad-57 to work
as a Masimo SET pulse oximeter without the Carboxyhemoglobin measurement. When
using one of these sensors, the front panel SpCO indicator will not be illuminated. See
Section 2, Rad-57 Front Panel Controls.
Select the appropriate patient cable to attach the LNOP or LNCS sensor to the device.
RED REUSABLE SENSORS
SpO
2
and pulse rate accuracy for the Red sensors is specified in the following table.
Weight Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
SENSOR Range No Motion Motion No Motion Motion Saturation Pulse Rate
Red
DC-I-DC
> 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3%
LNOP
®
REUSABLE SENSORS (LNOP sensors must be used in conjunction with Red PC cables)
Weight Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
SENSOR Range No Motion Motion No Motion Motion Saturation Pulse Rate
LNOP DC-I > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP DC-IP 10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP Y-I > 1 kg ± 2% ± 3% ± 3 bpm ± 5 bpm N/A N/A
LNOP TC-I > 30 kg ± 3.5% N/A ± 3 bpm N/A ± 3.5% ± 3 bpm
LNOP DC-195 > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP DCSC > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP TF-I > 30 kg ± 2% N/A ± 3 bpm N/A ± 2% ± 3 bpm
Note: The LNOP TF-I and TC-I sensors were not validated under motion conditions.
sensors & patient cables
LNOP
®
ADHESIVE SENSORS (LNOP sensors must be used in conjunction with Red PC cables)
Weight Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
SENSOR Range No Motion Motion No Motion Motion Saturation Pulse Rate
LNOP Adt > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP Adtx > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP Pdt 10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP Pdtx 10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP Neo < 10 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNOP NeoPt < 1 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNOP Neo-L
< 3 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
> 40 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP NeoPt-L < 1 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNOP Inf-L 3-20 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNOP
®
SPECIALTY SENSORS (LNOP sensors must be used in conjunction with Red PC cables)
Weight Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
SENSOR Range No Motion Motion No Motion Motion Saturation Pulse Rate
LNOP Blue 3-20 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNOP Hi Fi
Inf/Ped
3-10 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
10-30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNOP Hi Fi
Neo/Adult
< 3 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
> 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNCS
TM
REUSABLE SENSORS (LNCS sensors must be used in conjunction with Red LNC cables)
Weight Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
SENSOR Range No Motion Motion No Motion Motion Saturation Pulse Rate
LNCS DC-I > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNCS DC-IP 10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNCS TC-I > 30 kg ± 3.5% N/A ± 3 bpm N/A ± 3.5% ± 3 bpm
LNCS TF-I > 30 kg ± 2% N/A ± 3 bpm N/A ± 2% ± 3 bpm
Note: The LNCS TF-I and TC-I sensors were not validated under motion conditions.
LNCS
TM
ADHESIVE SENSORS (LNCS sensors must be used in conjunction with Red LNC cables)
Weight Saturation Accuracy Pulse Rate Accuracy Low Perfusion Accuracy
SENSOR Range No Motion Motion No Motion Motion Saturation Pulse Rate
LNCS Adtx > 30 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNCS Pdtx 10 - 50 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNCS Inf-L 3-20 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 2% ± 3 bpm
LNCS Neo-L
< 3 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
> 40 kg ± 2% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
LNCS NeoPt-L < 1 kg ± 3% ± 3% ± 3 bpm ± 5 bpm ± 3% ± 3 bpm
sensors & patient cables