Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual 8-1
8
sensors & patient cables
Introduction
This section covers the use and cleaning of Masimo sensors and patient cables.
Before use of any sensor, carefully read the sensor’s Directions for Use.
Use only Masimo oximetry sensors for SpO
2
and SpCO measurements. Other oxygen
transducers or sensors may cause improper Rad-57 Pulse CO-Oximeter performance.
Tissue damage can be caused by incorrect application or use of a sensor, for example by
wrapping the sensor too tightly. Inspect the sensor site as directed in the sensor Directions for
Use to ensure skin integrity and correct positioning and adhesion of the sensor.
CAUTIONS:
DO NOT USE DAMAGED SENSORS OR PATIENT CABLES. DO NOT USE A
SENSOR OR PATIENT CABLE WITH EXPOSED OPTICAL OR ELECTRICAL
COMPONENTS.
DO NOT IMMERSE THE SENSOR OR PATIENT CABLE IN WATER, SOLVENTS,
OR CLEANING SOLUTIONS (THE SENSORS AND CONNECTORS ARE NOT
WATERPROOF).
DO NOT STERILIZE SENSORS OR PATIENT CABLES BY IRRADIATION,
STEAM, AUTOCLAVE OR ETHYLENE OXIDE. SEE THE CLEANING
INSTRUCTIONS IN THE DIRECTIONS FOR USE FOR REUSABLE MASIMO
SENSORS.
DO NOT REPROCESS ANY MASIMO SET PATIENT CABLES OR SINGLE USE
SENSORS.
ALL SENSORS AND CABLES ARE DESIGNED FOR USE WITH SPECIFIC
MONITORS. VERIFY THE COMPATIBILIY OF THE MONITOR, CABLE AND
SENSOR BEFORE USE, OTHERWISE PATIENT INJURY CAN RESULT.
SELECTING A SENSOR
When selecting a sensor, consider, the patient’s weight, the adequacy of perfusion, the
available sensor sites, and the duration of monitoring. For more information refer to the
following table or contact your Sales Representative. Use only Masimo sensors and sensor
cables. Select an appropriate sensor, apply it as directed, and observe all warnings and
cautions presented in the Directions for Use accompanying the sensor.
High ambient light sources such as surgical lights (especially those with a xenon light
source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight can
interfere with the performance of the sensor. To prevent interference from ambient light,
ensure that the sensor is properly applied, and cover the sensor site with opaque material,
if required. Failure to take this precaution in high ambient light conditions may result in
inaccurate measurements.
SENSOR APPLICATION SITE
Unless indicated otherwise in the directions for use, reposition reusable sensors at least
every 4 hours and for adhesive sensors inspect the site at least every 8 hours or sooner. If
indicated by circulatory condition or skin integrity, reapply to a different monitoring site.