7-2 Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual
7
Rad-57 Signal Extraction CO-Pulse Oximeter Operator’s Manual 7-3
7
specifications specifications
1
Masimo Rainbow SET technology with LNOP, LNOPv and LNCS sensors has been validated for no motion
accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies in the range of 70-100%
SpO
2
against a laboratory CO-Oximeter and ECG monitor. 1% was added to account for the properties of
fetal hemoglobin. This variation equals plus or minus one standard deviation which encompasses 68% of the
population.
2
Masimo Rainbow SET technology with LNOP, LNOPv and LNCS sensors has been validated for motion accuracy
in human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping
motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude
of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO
2
against a laboratory co-oximeter and ECG
monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
3
Masimo Rainbow SET technology has been validated for low perfusion accuracy in bench top testing against a
Biotek Index 2 simulator and Masimo’s simulator with signal strengths of greater than 0.02% and a % transmission
of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard
deviation which encompasses 68% of the population.
4
This represents approximate run time at lowest indicator brightness, using a new, fully charged battery.
5
If alkaline batteries are to be stored for extended periods of time, it is recommended that they be stored between
-0°C to +30°C, and at a relative humidity less than 85%. If stored for a prolonged period at environmental conditions
beyond these limits, overall battery capacity may be diminished, and lifetime of the batteries may be shortened.
6
With FastSat the averaging time is dependent on the input signal. For the 2 and 4 second settings the averaging
time may range from 2-4 and 4-6 seconds, respectively.
7
SpCO accuracy has been validated on healthy adult volunteers against a laboratory co-oximeter. This variation
equals plus or minus one standard deviation which encompasses 68% of the population. The SpCO accuracy has
not been validated under motion conditions.
ELECTRICAL
Batteries
Type: 4 “AA” Alkaline
5
Capacity: over 8 hours
4
Isolation: No external power or ground connection, internally powered only.
ENVIRONMENTAL
Operating Temperature: 41°F to 104°F (5°C to 40°C)
Storage Temperature: -40°F to 158s°F (-40°C to +70°C)
5
Operating Humidity: 5% to 95%, non-condensing
Operating Altitude: 500 mbar to 1060 mbar pressure,
-1000 ft to 18,000 ft (-304 m to 5,486 m)
PHYSICAL CHARACTERISTICS
Dimensions: 6.2” x 3.0” x 1.4” (15.8 cm x 7.6cm x 3.6 cm)
Weight: 13oz. (0.32 kg)
SpO
2
Modes
Averaging mode: 2, 4, 8,10, 12, 14 or 16 seconds
6
Sensitivity: Normal, Maximum and APOD
Alarms
Audible and visual alarms for high low saturation and pulse rate
(SpO
2
range 1-100%, pulse rate range 25-240 bpm and SpCO %5 - %50)
Sensor condition, system failure and low battery alarms
High Priority: 571 Hz tone, 5 pulse burst, pulse spacing:
0.250s, 0.250s, 0.500s, 0.250s, repeat time:10s
Low Priority: 500Hz tone, 1 pulse, repeat time: 5s
Display/Indicators
Data display: %SpO
2
, %SpCO, SpCO bar, pulse rate, alarm status, alarm silenced status,
Perfusion Index (PI) / pleth bar, Low Signal IQ, battery status and SpCO connected.
Type: LED
Display update rate 1 second
Compliance
EMC Compliance: EN60601-1-2, Class B
Equipment Classification: IEC 60601-1-1
Type of Protection: Internally powered (on battery power)
Degree of Protection-Patient Cable: Type BF-Applied Part
Mode of Operation: Continuous