8-4 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
8
Masimo Sensors (continued)
LNOP
®
Specialty Sensors
LNOP sensors must be used in conjunction with Red PC cables.
• Newborn Infant/Pediatric
• Newborn Neonatal
• Trauma
• Blue
M-LNCS™/LNCS
®
Reusable Sensors
LNCS sensors must be used in conjunction with LNC cables.
• DCI
• DCIP
• YI
• TC-I
• TF-I
M-LNCS™/LNCS
®
Adhesive Sensors
LNCS sensors must be used in conjunction with Red LNC cables.
• Adtx
• Pdtx
• Inf
• Neo
• NeoPt
• NeoPt-500
M-LNCS™/LNCS
®
Specialty Sensors
LNCS specialty sensors must be used in conjunction with Red LNC cables.
• Newborn Infant/Pediatric
• Newborn Neonatal
• Trauma
SENSOR ACCURACY
See Sensor Directions for Use (DFU) for sensor accuracy specifications for: SpO
2
, SpMet,
SpCO, SpHb, and pulse rate.
CLEANING AND REUSE OF MASIMO REUSABLE SENSORS AND CABLES
Reusable sensors and patient cables can be cleaned per the following procedure:
1. Remove the sensor from the patient.
2. Disconnect the sensor from the patient cable.
3. Disconnect the patient cable from the monitor.
4. Wipe the entire sensor and/or patient cable clean with a 70% isopropyl alcohol pad.
5. Allow to air dry thoroughly before returning it to operation.
CAUTION: CAREFULLY ROUTE PATIENT CABLES TO REDUCE THE POSSIBILITY OF
PATIENT ENTANGLEMENT OR STRANGULATION.