4-32 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4
6. Name the file.
7. Save the file to the appropriate folder. Radical-7 trend information will consist of Date,
Time, SpO2, Pulse Rate and Perfusion Index. Trend information will include Pleth Variability
Index, SpCO, SpMet and/or SpHb, if the parameters are installed.
D) Disconnecting the Cable
1. Disconnect the USB-Serial or Serial-Serial cable from the instrument and PC. (See the
specific cable's Directions for Use for instructions on attaching the patient cable to the
instrument.)
ERASING TREND MEMORY
1. Ensure that the Radical-7 is in the Traditional User Interface.
2. Press
to access the trend menu.
3. Press
to enter the Trend menu.
4. Press
two (2) times to access the "Clear Trend" button.
5. Press
to delete Radical-7 trend data.
The Radical-7 continuously trends data. When performing a new study and gathering data on a
new patient, it is highly recommended the "clear function" be utilized in order for the results to be
separate. Turning the Radical-7 off will not erase the trend data.
TREND DATA FORMAT
After a successful download of the trend data, a .out file will be created containing the trend-dump
information in ASCII delimited format. The format is defined in the following table.
PARAMETER SPECIFICATION
Date
MM\DD\YY
Time
HH:MM:SS
Installed Parameter/
Measurement
Numeric value (see the display ranges in the Factory and User Configurable
Default Settings table located at the beginning of this section)
Exception
Messages
The exceptions are displayed as a 3 digit, ASCII encoded, hexadecimal
value. The binary bits of the hexadecimal value are encoded as follows:
000 = Normal operation; no exceptions
001 = No Sensor
002 = Defective Sensor
004 = Low Perfusion
008 = Pulse Search
010 = Interference
020 = Sensor Off
040 = Ambient Light
080 = Unrecognized Sensor
100 = reserved
200 = reserved
400 = Low Signal IQ
800 =
Masimo SET. This flag means the algorithm is running in full SET mode. It requires
a SET sensor and needs to acquire some clean data for this flag to be set
Operation