4-20 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4
Operation
Clock
MENU ITEMS DESCRIPTION
TIME
Set the time - hour, minutes and seconds - in 24 hour format.
TIME FORMAT
Set the format of the time display as it will be shown on the front panel.
Available options are 12 hour (default) and 24 hour display.
DAY
Set the numerical day.
MONTH
Set the numerical month.
YEAR
Set the numerical year.
DAY FORMAT
Set the format of the date display as it will be shown on the front panel.
Available options are mm/dd/yy (default) and dd/mm/yy.
DISPLAY CLOCK
Set to display the date and time on the front panel by selecting YES or NO.
About
This displays the copyright and software versions of the Handheld and Docking Station.
MENU ITEMS DESCRIPTION
MORE
LINE FREQENCY - Allows the user to view the current Line Frequency setting
without a password.
SENSOR TIME - Allows the user to view the amount of sensor time
remaining for SpHb Rainbow Reusable sensors.
NAC - Available battery life
LMD - Battery capacity
CPI - Number of charge cycles
Config
Config MENU ITEMS DESCRIPTION
ENTER PASSWORD
See Section 4,
Password Operation
Line Frequency
Set to 50 or 60 Hz.
Set to match regional power line frequency to allow
for cancellation of noise introduced by fl uorescent
lights and other sources. Factory default is 60 Hz.
Rapid Desat
Limit
Set to Off, -5 or -10.
When set to -5% or -10%, this feature overrides
the Alarm Delay (audible alarms silenced) and the
instrument will alarm if SpO
2
rapidly falls to 5% or
10% below the low alarm limit. Factory default is -5%.
Display
Measurement
Set to SpOC or None.
Display Measurement must be set to SpOC for
SpOC to show on the screen and in the Trend Setup
screen. Factory Default is SpOC.
SpHb UOM
Set to g/dL or mmol/L
Display total hemoglobin (SpHb) Unit of Measurement
data as g/dL (grams per deciliter) or mmol/L (milimoles
per liter). Factory default is g/dL.