Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual 7-3
7
Specifications
Sensor condition, system failure and low battery alarms
High Priority:
571 Hz tone, 5 pulse burst, pulse spacing: 0.250s,
0.250s, 0.500s, 0.250s, repeat time:10s
Medium Priority:
550 Hz tone, 3 pulse burst, pulse spacing: 0.375s,
0.375s, repeat time: 7s
Low Priority: 500 Hz tone, 1 pulse burst, repeat time: 5s
Alarm Muted reminder:
500Hz tone, 2 pulse burst, pulse spacing 0.375s,
repeat time: 3min.
Alarm Volume:
High Priority: 70 dB (min),
Medium Priority: 70 dB (min),
Low Priority: 45 dB (min)
Display/Indicators
Data display: %SpO
2
, %SpCO, %SpMet, SpHb g/dl, SpOC
ml/dl, PVI, pulse rate, plethysmographic
waveform, alarm status, trends, status messages, Signal IQ, perfusion index, APOD and
FastSat
Display update rate: 1 second
Response Time: < 20 second delay
Type: Backlit Active Matrix TFT LCD
Pixels: 480 x 272 dots
Dot Pitch: 0.25 mm
Output Interface
SatShare (RDS-1, RDS-1B)
Serial RS-232 (RDS-1, RDS-1B, RDS-3)
Nurse Call/Analog Output (RDS-1, RDS-1B, RDS-3)
Philips Vuelink, Spacelabs Universal Flexport, RadNet, Patient SafetyNet (RDS-1, RDS-1B, RDS-3)
Compliance
EMC Compliance: EN60601-1-2, Class B
Equipment Classification: IEC 60601-1 / UL 60601-1
Type of Protection Class 1 (on AC power), Internally powered (on battery power)
Degree of Protection-Patient Cable: Type BF-Applied Part
Degree of Protection-SatShare Cable: Type CF-Applied Part
Mode of Operation: Continuous
1 SpO
2
,
SpCO and SpMet accuracy was determined by testing on healthy adult volunteers in the range 60% - 100% SpO
2
,
0% - 40%
SpCO and 0% - 15% SpMet against a laboratory CO-Oximeter. SpO
2
and SpMet accuracy was determined on 16 neonatal NICU
patients ranging in age from 7 to 135 days old and weighting between 0.5 and 4.25 kgs. Seventy-nine (79) data samples were collected
over a range of 70 - 100% SaO
2
and 0.5 - 2.5% HbMet with a resultant accuracy of 2.9% SpO
2
and 0.9% SpMet. Contact Masimo
for testing speciations.
2 The Masimo Rainbow SET technology with Masimo sensors has been validated for no motion accuracy in human blood studies on
healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of 70-100% SpO
2
against a laboratory CO-Oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one
standard deviation encompasses 68% of the population weight.
3
The Masimo Rainbow SET technology with Masimo sensors has been validated for motion accuracy in human blood studies on
healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies while performing rubbing and
tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm
in induced hypoxia studies in the range of 70-100% SpO
2
against a laboratory CO-Oximeter and ECG monitor. This variation equals
plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
Radical-7 Specifications (continued)