4-16 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4
Operation
MENU ITEMS DESCRIPTION
SILENCE
This menu allows the user to set the alarm silence period for SpO
2
,
SpMet, SpCO, SpHb, PI, PVI and pulse rate. An alarm is silenced by
pressing the Alarm Silence button on the front panel.
30, 60, 90,120 SECONDS
The alarm silence can be set for these durations. As an indicator that
the alarm system is silenced, the Alarm Status Indicator is shown as a
bell with a slash through it. A timer is shown next to the bell indicating
the remaining alarm silence duration.
NOTE: The alarm silence period is reset to 120 seconds (or 90
seconds in neonatal mode) upon power cycle, except when the
Radical-7 is set to operate in the Home mode.
ALL MUTE
All patient alarm conditions are silenced. Only system alarms will be
indicated by an audible alarm. As an indicator that the system is set to All
Mute, the Alarm Status Indicator is shown as a bell with a slash through it.
ALL MUTE WITH AUDIBLE REMINDER
All patient alarm conditions are silenced. Only system alarms will be
indicated by an audible alarm. As a reminder, a single audible alarm will
occur every three minutes. As an indicator that the system is set to All
Mute, the Alarm Status Indicator is shown as a bell with a slash through it.
WARNING: IF AN ALARM CONDITION OCCURS WHILE THE ALARM SILENCE
PERIOD IS SET TO ALL MUTE, THE ONLY ALARM INDICATIONS
WILL BE VISUAL DISPLAYS AND SYMBOLS RELATED TO THE
ALARM CONDITION. NO ALARM TONE WILL SOUND.
NOTE: With the Handheld instrument, if there is a loss of power for any length of time, check
the alarm settings. Generally, the alarm limit settings will be set back to the User set defaults
if the user previously selected YES in the Save Last option under the Display menu. If the user
has not utilized this option, then they will be set back to the factory defaults.
MENU ITEMS DESCRIPTION
VOLUME
This menu allows the user to set the alarm volume. Four levels are
available: level 1 being the softest and level 4 being the loudest. The
instrument retains the Alarm Volume setting upon a power cycle.
NOTE: For home use, set the alarm level to level 4.