4-22 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
4
MENU ITEMS DESCRIPTION
INTERFACE
ALARMS
SpO
2
/BPM
During SatShare, Patient SafetyNet, Philips Vuelink, Spacelabs Flexport
and RadNet operation, the audible alarms for SpO
2
can be enabled or
disabled by selecting Yes or No.
All other parameters/measurements will be actively monitored by the instrument.
The Rainbow parameters/measurements cannot be turned off with this
feature.
Yes setting: allows SpO
2
and pulse rate audible alarms at the
Radical-7 and the interfaced system.
No setting: mutes the SpO
2
and pulse rate audible alarms at the
Radical-7 while allowing SpO
2
and pulse rate audible
alarm alerts at the interfaced system. The “No” setting
prevents both systems (Radical-7 and interfaced system)
from producing audible alarms at the same time.
NOTE: The Radical-7 reverts to Interface Alarm SpO
2
/BPM “Yes” during
power interruptions or when the SatShare connection is lost or the
instrument becomes separated from the docking station. This ensures that
the Radical-7 provides audible alarms for SpO
2
and pulse rate when the
connection to the interfaced system becomes compromised.
SATSHARE
NUMBERS
During SatShare operation the saturation and pulse rate measurements can be
displayed on the Radical-7 by selecting a SatShare Numbers setting of Yes.
POWER SAVE
Select Yes to maximize battery-operating time of the Radical-7 while
powered by the Handheld battery or optional Docking Station battery.
Selecting Yes will disable Docking Station functions such as SatShare,
Serial and Analog output. Selecting No will activate these Docking
Station functions while operating on battery power. (While operating in
the Power Save mode, a power cycle of the Radical-7 may be required to
activate the Docking Station again after it has been disabled.)
NOTE: Handheld devices with software version 7.6 or greater will
automatically activate the power save mode when the handheld device
is docked into a docking station that is not connected to AC power.
SMART TONE
Select Yes to activate the SmartTone function. This will allow the audible
pulse to continue to beep when the pleth graph shows signs of motion.
Select NO to turn off SmartTone.
PVI
Select Yes to display the PVI measurement. The PVI will display
numerically on the main screen and also will allow to the user to change
the maximum and minimum PVI settings in the Trend Setup menu.
Select NO to deselect the measurement.
SpHb AVERAGING
SpHb Averaging can be set to Long, Medium, or Short.
PI AVERAGING
The signal averaging algorithm can be set to short or long. The default
is short.