Masimo Radical Signal Extraction Manuale dell'Operatore Pagina 96

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 95
7-6 Radical-7 Signal Extraction Pulse CO-Oximeter Operator’s Manual
7
PIN SIGNAL NAME
7
Nurse Call (Normally Closed)
8
Ground
9
Analog 1
10
Ground
11
Ground
12
Nurse Call – Common
13
Ground
14
Ground
15
Analog 2
ANALOG OUTPUT
The Radical-7 Pulse CO-Oximeter can interface with various analog recording instruments
and/or strip chart recorders through its Analog Output connector located on the back of
the Docking Station. Depending on the configuration of the Output menu, the following
parameters are output continuously on the Analog 1 and Analog 2 channels:
SpO
2
Pulse rate
Plethysmographic waveform
Signal IQ
The output signals vary from approximately 0 to 1 volt in a linear fashion.
NOTE: The actual Analog 1 and Analog 2 output voltage that are generated may not exactly
range between 0.0V to 1.0V. A variance of ± 40 mV is acceptable.
CALIBRATION
For measurement instrument calibration purposes, the analog output signals can be set to
either 0 Volts or 1 Volt in the menu system under Output/Analog Output Mode. Calibrate your
analog recording system to those levels before use.
NURSE CALL
The nurse call feature is available when Radical-7 Pulse CO-Oximeter is operating in its
standalone configuration. The nurse call feature on the Radical-7 Pulse CO-Oximeter is
based on the relay closing or opening depending on alarm, Low Signal IQ events or both.
For maximum flexibility, either normally open (pin 6) or normally closed (pin 7) signals are
available. Only qualified personnel should connect one of these two signals and common (pin
12) to a hospital’s nurse call system. During an alarm condition, or a Low Signal IQ event,
depending on the configuration of the output menu, the normally open pin will be connected
to the common pin and the normally closed will be disconnected. In addition, the nurse call
polarity can be inverted to accommodate various nurse call station requirements.
Specifications